AFE914ABGA64
-
Ultralow-power, 4-channel 16-bit ADC biopotential ECG analog front end
Product details
-
Parameters
-
Characteristics
-
Package
-
Description
-
-
12-lead and 18-lead ECG medical equipment
-
4-channel, 16-bit, SPI communication
-
-20°C to+70°C
-
-
-
• 4 low-noise programmable gain amplifiers (PGA) and 4 high-resolution analog-to-digital converters (ADC)
-
Low power consumption
0.33mW per channel
-
• Input-referred noise
12μVPP (150Hz BW,G = 6)
-
• Input bias current
50pA
-
• Data rate
125PS to 8kSPS
-
• Common Mode Rejection Ratio
-110dB
-
• Programmable gain
1、2、3、4、6、8、12 or 24
-
• Supports systems compliant with AAMI EC11, EC13, IEC 60601-1, IEC 60601-2-27, and IEC 60601-2-51 standards
-
• Unipolar or bipolar power supply
-
AVDD = 2.7V to 5.5V
-
DVDD = 1.65V to 3.6V
-
• Integrated right leg drive amplifier, lead-off detection, Wilson central terminal, pace detection, and test signal
-
• Built-in AC lead-off excitation
-
• Built-in oscillator and reference
-
• Serial Peripheral Interface (SPI™)-compatible serial port, daisy-chain
-
-
-
QFP64 10mm x 10mm
-
BGA64 8mm x 8mm
-
-
The AFE914 is a series of multi-channel, synchronous sampling 16-bit Δ-Σ analog-to-digital converters (ADCs) featuring a built-in programmable gain amplifier (PGA), internal reference, and an on-board oscillator. The AFE91x encompasses all the necessary functionalities typically required in medical electrocardiogram (ECG) and electroencephalogram applications. With its high level of integration and exceptional performance, the AFE91x enables the development of scalable medical instrument systems with significantly reduced size, power consumption, and overall cost.
Each channel of the AFE91x includes a flexible input multiplexer (mux) that can independently connect to internally generated signals for testing, temperature, and lead-off detection. Additionally, any configured input channel can generate a right leg drive (RLD) output signal. Lead-off detection can be implemented within the device using pull-up/pull-down resistors or excitation current sinks/sources. Three integrated amplifiers are used to generate the Wilson Central Terminal (WCT) required for standard 12-lead ECG. Multiple AFE91x devices can be connected in a daisy-chain configuration in high channel count systems.
The device is housed in a compact 8mm × 8mm, 64-ball BGA package. The BGA version is rated for an industrial temperature range of 0°C to +70°C.
Related applications
Related products
-
ADX920QAQFN32
Automotive-grade, low-noise PGA and 24-bit high-precision ADC analog ECG front end
-
ADX921QAQFN32
Automotive-grade, two low-noise PGAs and two 24-bit high-precision ADC analog ECG front end
-
ADX922QAQFN32
Automotive-grade integrated respiration impedance, two low-noise PGAs, and two 24-bit high-precision ADCs for an analog ECG front-end.