ADX923AQFN32
- 
                                        Integrated respiration impedance, three low-noise PGAs, and two 24-bit high-precision ADCs for analog ECG front-end. 
Product details
- 
                                            Parameters 
- 
                                            Characteristics 
- 
                                            Package 
- 
                                            Description 
- 
                                            - 
                                                        Integrated respiration impedance, three PGAs, two 24-bit high-precision ADCs, SPI communication, and a sampling rate of 16Ksps. 
- 
                                                        -40°C to+85°C 
 
- 
                                                        
- 
                                            - 
                                                        •Three low-noise PGA (ADX923) and two high-resolution ADCs (ADX922/3). 
- 
                                                        • Input-referred noise. 8.6μVPP (G = 6) 
- 
                                                        • Input bias current 100pA 
- 
                                                        • Data rate 125SPS 至 8kSPS 
- 
                                                        • Common Mode Rejection Ratio 117dB 
- 
                                                        • Programmable gain 1, 2, 3, 4, 6, 8, or 12 
- 
                                                        Analog power supply 2.7V to 5.5V 
- 
                                                        digital power supply 1.65V to 3.6V 
- 
                                                        • Built-in right leg drive amplifier, lead-off detection, test signal. 
- 
                                                        • Integrated respiratory impedance measurement (ADX922/3). 
- 
                                                        • Electrode Rotation Mode (ERM) enables time-division multiplexing for LEAD-I, LEAD-II, and LEAD-III measurements (ADX923). 
- 
                                                        • Integrated simulation PACE measurement (ADX921/2/3) 
- 
                                                        High speed mode (ADX921/2/3) 
- 
                                                        • Built-in Oscillator and Reference 
- 
                                                        • Power supply monitor, internal temperature sensor, OSC failure detection, over-range detection. 
- 
                                                        • Flexible power-down and standby modes 
- 
                                                        • SPI™ compatible serial interface with timeout reset functionality 
- 
                                                        • Communication with cyclic redundancy check (CRC) and checksum 
- 
                                                        • Supports ADC data FIFO to reduce polling time for external devices and conserve power 
- 
                                                        • Functional safety 
- 
                                                        • Supports systems compliant with IEC 60601-2-47 standard • Operating temperature range –40°C to 85°C 
 
- 
                                                        
- 
                                            - 
                                                        QFN32 4mm x 4mm 
 
- 
                                                        
- 
                                            The ADX920/1/2/3 devices are multi-channel, simultaneous sampling, 24-bit, Delta-Sigma (ΔΣ) analog-to-digital converters (ADCs) that feature a programmable gain amplifier (PGA), an internal reference, and an on-board oscillator. These devices integrate all the functionalities typically required for portable, low-power medical electrocardiogram (ECG) applications, as well as for fitness and exercise monitoring systems. With their high level of integration and excellent performance, the ADX920/1/2/3 devices enable the creation of scalable medical instrument systems with significantly reduced size, power consumption, and overall cost. Each channel of the ADX920/1/2/3 devices includes a flexible input multiplexer that can independently connect to internally generated signals for testing, temperature measurement, and lead-off detection. Moreover, any input channel configuration can be configured to derive a right-leg drive (RLD) output signal. The ADX920/1/2/3 devices operate at data rates up to 8 kSPS. Lead-off detection can be accomplished internally using the device's current sink or source excitation capabilities. The ADX922/3 models additionally feature fully integrated respiratory impedance measurement functionality. The specified operating temperature range is from –40°C to 85°C. 
Related applications
Related products
- 
                                        
                                            ADX920QAQFN32Automotive-grade, low-noise PGA and 24-bit high-precision ADC analog ECG front end 
- 
                                        
                                            ADX921QAQFN32Automotive-grade, two low-noise PGAs and two 24-bit high-precision ADC analog ECG front end 
- 
                                        
                                            ADX922QAQFN32Automotive-grade integrated respiration impedance, two low-noise PGAs, and two 24-bit high-precision ADCs for an analog ECG front-end. 